When an online chlorine reading differs from a reference test

Investigate a difference between an online chlorine sensor and a DPD reference test. Check the measured residual, sample route and timing before deciding whether to recalibrate.

Halogen D20 mounted inside a light service room overlooking a treatment plant

When an online chlorine reading differs from a reference test, first check that both methods measure the same disinfectant and compare water from the same point and time. Verify sample flow and the reference method before investigating calibration.

Check that both methods measure the same residual

Confirm the analyser channel, units and test method in use. A free-chlorine reading and a total-chlorine result can differ even when both instruments are working as intended. For chloraminated water, confirm whether the question calls for total chlorine or a species-specific measurement, and choose a reference method suited to that question.

Establish where the sample came from and when

Record the sampling point and time alongside the online trend. If the analyser is fed through a sample line, a change at the take-off reaches the instrument after a delay. A grab sample taken elsewhere or at a different time may represent a different parcel of water. Use the sampling point and procedure in the site plan.

Look for changes in the sample supply

Check the recorded valve position, flow indication and any change to the water path since the last satisfactory check. Follow the site’s safe access and isolation procedures. PMA’s published MP5 field account describes a low reading investigated by checking sample supply and confirming the restored residual with a reference measurement.

Check the reference test and operating conditions

Use the operator’s approved reference procedure and check the equipment and reagents are suitable for the expected residual. Record conditions such as pH and temperature where they matter to the selected instrument. Capture any alarm or diagnostic state and the reported channel at the time of the comparison.

Make the decision using the approved procedure

Repeat the comparison under suitable conditions if the initial checks reveal a timing or sampling problem. If a repeatable difference remains, use the applicable instrument manual and site criteria to decide whether calibration, servicing or an investigation is required. Escalate unresolved readings under the site’s operating procedures.

Keep a record the next operator can use

Keep the sample point, residual species, test method, timestamps, both readings, operating conditions and any action together. Record the follow-up result so that others can distinguish a restored sample supply from a change made to the instrument.

Your questions answered.

Does a difference mean the online sensor needs calibration?

Check the residual being measured, sampling point and timing, sample supply and reference procedure first. If a repeatable difference remains, follow the exact model’s instructions and the site’s acceptance criteria.

Is there one acceptable difference for every installation?

Use the site’s acceptance criteria and the relevant instrument and reference method documentation. The comparison depends on the residual, measurement range, methods and conditions.

Technical references.

Read alongside the current documentation for your exact sensor and installation.

PMA’s MP5 field-trial account

Read the sample-supply investigation and reference check from one installation.

Read reference

Halogen free chlorine sensor documentation

Identify the applicable model and documentation before changing calibration or service settings.

Read reference

Find the right fit for your water.

Tell us about your disinfectant, water chemistry and installation. PMA can help you work through the sensor, controller and integration requirements.

Start a conversation.

01484 843 708sales@pma.uk.com

Brook Mills House, Carr Lane
Slaithwaite, Huddersfield, HD7 5BG